Near and Intermediate Visual Acuity in Clinical Studies with Multifocal Intraocular Lenses

DCIVA and DCNVA are the standardized abbreviations for the measurement of visual acuity at intermediate and near distances when the patient is corrected with best spectacle correction for an infinite distance.[1] CIVA and CNVA are the abbreviations for visual acuity at intermediate and near when the patient is corrected with best spectacle correction for each of these distances.[1]

Protocols for clinical studies should be designed to include the minimum number of tests that maximize the information obtained. Failure to understand the purpose of why certain visual performance measures are taken can lead to the development of clinical trial protocols with long duration and therefore unreliable data due to patient fatigue.

Many clinical studies include visual acuity with the best near and intermediate correction (CNVA and CIVA) with multifocal intraocular lenses, even these variables are included in the international standards of Clinical Studies[2]. In our opinion, these measures are a clear example of redundant measures that increase evaluation times providing little or no additional information.

Why is the information redundant?

To explain why we're going to base it on the defocus curve.  In the image you can see the relationship between the positions of the defocus curve and each of these abbreviations. Figure A shows how the patient with best infinity correction will have the peak of maximum vision at 0 D on the defocus curve. The abbreviations DCIVA and DCNVA correspond to the -1.50 D (67 cm) and -2.50 D (40 cm) curve positions.

CNVA and CIVA

When we seek to measure the CIVA and CNVA what we do is to place the addition that optimizes the vision of the patient at intermediate and near distances, considering that the peak of best vision is at 0 D, placing an addition of +1.50 D would optimize the vision at 67 cm (Figure B) and with +2.50 D at 40 cm (Figure C), but the visual performance is the same as that obtained in the CDVA since we are measuring the visual performance provided by the same focus in different positions in a redundant way after shifting the defocus curve.

Why do CDVA, CIVA and CNVA differ in clinical studies?

Although in theory CDVA, CIVA and CNVA should be the same in clinical studies, the means may differ for several reasons, including:

  1. A 5 cm error between the patient's position and the test when the desired position is 40 cm generates a 0.25 D defocus so for the CDVA and the CIVA/CNVA to coincide there should be perfect control of the presentation distance.
  2. Other sources of bias such as the magnifying effect of positive lenses could influence the results.
  3. Using different tests for Far, Intermediate and Near: Optotypes, Lighting, etc.

References:

  1. Dupps WJ, Kohnen T, Mamalis N, et al (2011) Standardized graphs and terms for refractive surgery results. J Refract Surg 27:7–10
  2. ISO-11979-7:2018 (2018) Ophthalmic implants - Intraocular lenses - Part 7: Clinical investigations
Do you prefer to learn with video tutorials?
Analysis of Results in Vision Sciences

Online Course: Analysis of Results in Refractive and Cataract Surgery

We have prepared a complete course that will help you to understand the standards for reporting the clinical results of your practice. This is a practical course, besides the theoretical knowledge, we are going to supply you with the "Refractive Analysis Toolbox" for Matlab that will help you to do all the analyses by yourself. Don't be afraid about Matlab because we have reduced the complexity of the calculations to specific functions that give you the results in a easy way.

Join Our Facebook Group

Access to our Facebook Group to keep you up to date with the latest news or to get support

Go now!

Terms & Conditions

Review the Terms & Conditions before to use our Apps and Courses

Review now!

Review Privacy Policy of this website

Review now!